Surgical shadowless lamp has been popularized in hospitals and clinics, surgical shadowless lamp is simple to operate, easy to use, in order to better play its advantages, we need to adjust the equipment correctly. If you do not know much about the debugging method, the manufacturer today will share with you the correct debugging method of the surgical shadowless lamp, I hope you can read it carefully, and look forward to your reading and attention.
1. Installation inspection
the main thing is to see whether the screws are all in place and tightened during the installation process, whether the various decorative covers have been covered, or whether there are other installation omissions.
2. Circuit inspection
This is a safety check point of surgical shadowless lamp, first of all, in the case of power failure, check whether there is a short circuit or open circuit phenomenon of the shadowless lamp; If not, check whether the shadowless lamp power supply is stable after powering on, and whether the transformer input voltage is stable and meets the requirements of the shadowless lamp.
3. Balance arm adjustment
When medical staff adjust the position of surgical shadowless lamp, they all need the balance arm system to bear the force, so it is necessary to check whether the balance arm can be adjusted to the angle required by the medical staff and whether it can be forced.
4. Joint flexibility
Because of the need to adjust the shadowless lamp angle, joint flexibility is also very important, mainly to adjust the joint damping screw, the standard stipulates that the force should be close to 20N or 5Nm when the tightness of damping adjustment is to push or rotate the joint in any position
5. Illumination depth
Because doctors are likely to need to look at the depth of the patient's wound during surgery, so the surgical shadowless lamp is required to have a good lighting depth, generally at a distance of 700-1400mm is better, about this, JoRadiant shadowless lamp fully meets the requirements.
6. Illuminance and color temperature inspection
This is the more important point of surgical shadowless lamp, good illumination and color temperature help doctors carefully observe the patient's wound tissue, distinguish organs, blood, etc., so close to the illuminance of sunlight and 4400-4600K color temperature is more suitable.
About the formal adjustment method of the surgical shadowless lamp will be briefly introduced to you here, I hope this section can help you solve some problems, if you have other problems to understand, then you can contact our customer service online, our customer service will help you in time, thank you!