Here are some Common faults and troubleshooting methods of high-frequency electrosurgical unit:
Possible cause: The power setting is too high
Elimination method: refer to the adjustment table in the electrode box, and set the power level according to the cutting situation and the selected electrode
Possible reason1: Mains power is not connected. Remedy: Check the mains outlet
Possible reason2: Insurance burnt. Remedy: Replace the insurance
Possible reason3: The foot switch is wrongly connected. Remedy: Check the foot brake
Possible cause: The power setting is too low
Elimination method: increase the power, step by step, until the cutting occurs, it is not necessary to be too high
Possible cause: The neutral plate is not connected
Remedy: Check the neutral plate connection
Possible cause: wrong neutral plate placement
Remedy: Check that the "PATIENT SIDE" side of the neutral plate is facing the patient Make sure the surface of the neutral plate is in full contact with the patient (through a thin layer of clothing), such as at the shoulder blades
Possible cause: The electrode is too dirty
Remedy: Turn off the switch of the electrosurgical knife (in the "O" state), wipe the electrode with a tissue dipped in oxygenated water
Possible cause: The electrode moves too fast
Solution: Slow down the operation